УДК: 535.233.4
Calibration of blackbody simulators for thermal imaging and pyrometry
Full text «Opticheskii Zhurnal»
Full text on elibrary.ru
Publication in Journal of Optical Technology
Алешко Е.И., Гараева А.И., Курт В.И. Калибровка моделей абсолютно черных тел для тепловидения и пирометрии // Оптический журнал. 2017. Т. 84. № 3. С. 47–48.
Aleshko E.I., Garaeva A.I., Kurt V.I. Calibration of blackbody simulators for thermal imaging and pyrometry [in Russian] // Opticheskii Zhurnal. 2017. V. 84. № 3. P. 47–48.
E. I. Aleshko, A. I. Garaeva, and V. I. Kurt, "Calibration of blackbody simulators for thermal imaging and pyrometry," Journal of Optical Technology. 84(3), 195-196 (2017). https://doi.org/10.1364/JOT.84.000195
This paper discusses problems related to the performance verification and calibration of blackbody simulators, which are used for determining the characteristics of optoelectronic devices operated in various spectral regions. It is shown that the brightness temperatures of a blackbody simulator differ in various spectral ranges at the same thermodynamic temperature. The averaged corrections to the thermometer readings of the radiator being certified and the correction magnitudes calculated at a specific radiation wavelength are compared.
thermal imaging device, blackbody simulator, calibration, blackbody simulator spectral range, corrections to thermometer readings of calibrated blackbody simulator
OCIS codes: 120.5240, 120.3940
References:1. GOST 8.558−2009 GSI, “State verification procedure for temperature measurement devices,” paragraph 4.3.
2. GOST R 8.566−2012 GSI, “Blackbody radiators. A standard procedure for their verification and calibration.”
3. GOST R 8.619−2006 GSI, “Thermal imaging measurement instruments. A standard procedure for their verification.”