ITMO
ru/ ru

ISSN: 1023-5086

ru/

ISSN: 1023-5086

Scientific and technical

Opticheskii Zhurnal

A full-text English translation of the journal is published by Optica Publishing Group under the title “Journal of Optical Technology”

Article submission Подать статью
Больше информации Back

DOI: 10.17586/1023-5086-2018-85-04-41-48

УДК: 681.785.422

Measurement traceability of total radiance factors for white fluorescent materials employing incandescent-lamp Zeiss leucometer photometers and with modern spectrophotometers based on a D65 artificial illuminant

For Russian citation (Opticheskii Zhurnal):

Горицкий Е.И. Обеспечение единства измерений общих коэффициентов яркости белых флуоресцирующих материалов на фотометрах типа лейкометра Цейсса с лампой накаливания и современных спектрофотометрах с искусственным источником D65 // Оптический журнал. 2018. Т. 85. № 4. С. 41–48. http://doi.org/10.17586/1023-5086-2018-85-04-41-48

 

Goritskiy E.I. Measurement traceability of total radiance factors for white fluorescent materials employing incandescent-lamp Zeiss leucometer photometers and with modern spectrophotometers based on a D65 artificial illuminant [in Russian] // Opticheskii Zhurnal. 2018. V. 85. № 4. P. 41–48. http://doi.org/10.17586/1023-5086-2018-85-04-41-48

For citation (Journal of Optical Technology):

E. I. Goritskiĭ, "Measurement traceability of total radiance factors for white fluorescent materials employing incandescent-lamp Zeiss leucometer photometers and with modern spectrophotometers based on a D65 artificial illuminant," Journal of Optical Technology. 85(4), 222-228 (2018). https://doi.org/10.1364/JOT.85.000222

Abstract:

A discussion based on the theory of measurements using an upgraded Zeiss leucometer (0/d) in which the fraction of UV radiation from the incandescent lamp is controllable was used to obtain a photometric equation (appropriate for any modern leucometer-type photometer) relating the total reflectance factors measured using such an instrument to the total spectral radiance factors for white fluorescent materials measured using a modern spectrophotometer (d/0) with a D65 artificial illuminant at the leucometer effective wavelength. A review of the available experimental results using the resulting equation indicated that it was possible to obtain measurement traceability of the total spectral radiance factors for white fluorescent materials when employing Zeiss leucometer photometers with an incandescent lamp and by using modern spectrophotometers with a D65 artificial illuminant.

Keywords:

photometer, spectrophotometer, reflection coefficient, fluorescence, UV

OCIS codes: 120.0120, 260.0260

References:

1. R. McDonald, ed., Colour Physics for Industry (Society of Dyers and Colourists, Bradford, England, 1997; Logos, Moscow, 2002).
2. D. B. Judd and G. Wyszecki, Color in Business, Science, and Industry, 3rd ed. (Wiley, New York, 1975; Mir, Moscow, 1978).
3. E. I. Goritskiı˘, “The possibility of using an incandescent lamp instead of an artificial D65 source in modern photometers of the Zeiss leucometer type,” J. Opt. Technol. 81(3), 146–150 (2014) [Opt. Zh. 81(3), 44–50 (2014)].
4. GOST 30113-94 (ISO 2470-77), “Paper and Board: Method for Determination of Whiteness Factor,” Izdatelstvo Standartov Printing and Publishing House, Moscow, 1996.
5. E. I. Goritskiı˘, “Theory of measurements of the reflectances of light-scattering materials on simplified sphere spectrophotometers with zonal light filters,” Sov. J. Opt. Technol. 57(5), 267–270 (1990) [Opt. Mekh. Promst. (5), 3–7 (1990)].
6. E. I. Goritskiı˘, “Ensuring consistency in measurements of the reflectances of white light-diffusing materials on FO-1 photometers and SF-18 spectrophotometers,” Sov. J. Opt. Technol. 60(5), 349–353 (1993) [Opt. Zh. (5), 61–67 (1993)].
7. E. I. Goritskii, “Theory and practice of measurements of the reflectances of light-scattering materials based on simplified vapor spectrophotometers with zone filters,” Meas. Tech. 36(8), 881–887 (1993) [Izmer. Tekh. (8), 26–30 (1993)].